Rack Mount the Instrument
You can mount the instrument in a standard 19-inch rack cabinet using one of two optional kits, each of
which includes instructions and mounting hardware. You may also mount Another Keysight System II
instrument of the same height and width beside the instrument.
To prevent overheating, do not block airflow to or from the instrument. Allow enough
clearance at the rear, sides, and bottom of the instrument to permit adequate internal
air flow.
Remove the carrying handle and the front and rear bumpers before rack-mounting the instrument.
Removing the Handle and Bumpers
To remove the handle, rotate it to vertical and pull the ends outward.
To remove the rubber bumper, stretch a corner and then slide it off.
Front
Rear (bottom view)
Rack Mounting a Single Instrument
To rack mount a single instrument, order adapter kit 5063-9240.
Rack Mounting Instruments Side-by-Side
To rack mount two instruments side-by-side, order lock-link kit 5061- 8769 and flange kit 5063-9212. Be
sure to use the support rails in the rack cabinet.
Quick Start
44
Keysight True
volt
Series Operating and Service Guide