Performance Verification Tests
The performance verification tests are recommended as acceptance tests when you first receive the instru-
ment. The acceptance test results should be compared against the 90 day test limits. You should use the
24 hour test limits only for verification within 24 hours after performing the calibration adjustment pro-
cedure.
If a Fluke 5725A Calibrator is not available, current verification can be limited to 2A. This will limit the
span of verification on the 10A ranges to only 2A.
If the instrument fails performance verification, calibration adjustment or repair is required.
To reset the date used to determine the specifications versus time, a calibration with adjustment is
required.
The allowable errors for verification are the published specifications (see the instrument data sheet at
www.keysight.com/find/truevolt-doc
). The periods in the published specifications are since the last full
calibration adjustment.
Recommended Test Equipment
The recommended test equipment for the performance verification and calibration is listed below. If the
exact instrument is not available, substitute calibration standards of equivalent accuracy.
An alternate method is to use the Keysight 3458A 8½-digit Digital Multimeter to measure less accurate,
yet stable, sources. The output value measured from the source can be entered into the instrument as the
target calibration value.
Application
Recommended Equipment
Zero Calibration
Keysight 34172B
DC Voltage
Fluke 5720A
DC Current
Fluke 5720A
Resistance
Fluke 5720A
AC Voltage
Fluke 5720A
AC Current
Fluke 5720A
Frequency
Keysight 33500B Series Function/Arbitrary Waveform Generator
High Current
Fluke 5725A
Capacitance (Optional) SCA-1uF Capacitance Standard
Keysight True
volt
Series Operating and Service Guide
505
Performance Verification