Values for 34461A DMM
Error from Nominal
Input Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
Open
DC Current
100 µA
Q
± 0.02 µA
± 0.025 µA
1 mA
± 0.060 µA
10 mA
± 2 µA
100 mA
± 4 µA
± 5 µA
1 A
± 60 µA
± 100 µA
3 A
± 600 µA
10 A
± 1 mA
Short
DC Volts
100 mV
Q
± 3 µV
± 3.5 µV
1 V
± 6 µV
± 7 µV
10 V
± 40 µV
± 50 µV
100 V
± 600 µV
1000 V
± 6 mV
± 10 mV
Short 4-wire Ohms
100 Ω
± 3 mΩ
± 4 mΩ
1 kΩ
± 5 mΩ
± 10 mΩ
10 kΩ
Q
± 50 mΩ
± 100 mΩ
100 kΩ
± 500 mΩ
± 1 Ω
1 MΩ
± 10 Ω
10 MΩ
± 100 Ω
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Keysight True
volt
Series Operating and Service Guide
507
Performance Verification