DC Volts and DC Current Gain Verification
This procedure verifies DC volts and DC current gain.
1.
Make sure you have read
.
2.
Connect the calibrator to the input terminals. For the 34461A/65A/70A, use the front panel input ter-
minals and select the front input terminals with the
Front/Rear
switch.
3.
Configure each function and range in the order shown in the table below corresponding to the DMM
model number. Provide the input shown in the table.
4.
Make a measurement and return the result. Compare measurement results to the test limits shown
below. (Be certain to allow for appropriate source settling.)
Values for 34460A DMM
Input
Error from Nominal
Voltage Function
[1]
Range
Quick
Check
24 hours
90 days
1 year
2 years
100 mV
DC Volts
100 mV
± 10 µV
± 13.5 µV ± 15.5 µV
± 18 µV
-100 mV
1 V
1 V
± 39 µV
± 70 µV
± 90 µV
± 115 µV
-1 V
4 V
10 V
± 140 µV
± 250 µV
± 350 µV
± 450 µV
10 V
10 V
Q
± 290 µV
± 550 µV
± 800 µV
± 1.1 mV
-10 V
100 V
100 V
± 3.6 mV
± 7.1 mV
± 9.1 mV ± 11.6 mV
-100 V
Q
1000 V
1000 V
± 36 mV
± 75 mV
± 95 mV
± 120 mV
-500 V
± 21 mV ± 42.5 mV ± 52.5 mV ± 62.5 mV
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Values for 34460A DMM
Input
Error from Nominal
Current Function
[1]
Range
Quick
Check
24 hours
90 days
1 year
2 years
2 A
DC Current
3 A
± 4.2 mA
± 4.6 mA
± 4.6 mA
± 5.2 mA
1 A
1 A
Q
± 560 µA
± 900 µA
± 1.1 mA
± 1.3 mA
100 mA
100 mA
± 14 µA
± 35 µA
± 55 µA
± 65 µA
10 mA
10 mA
Q
± 2.7 µA
± 5 µA
± 7 µA
± 8 µA
1 mA
1 mA
± 0.13 µA ± 0.36 µA
± 0.56 µA
± 0.66 µA
100 µA
100 µA
± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA
[1]Select 10 PLC.
Q
: Quick performance verification test point.
Performance Verification
510
Keysight True
volt