ACV Values for 34461A DMM
Input
Error from Nominal
Vrms
Frequency Range
Filter
Quick
Check
24 hours 90 days
1 year
2 years
100 mV
1 kHz
100 mV 200 Hz
± 60 µV
± 80 µV
± 90 µV
± 100 µV
50 kHz
± 140 µV ± 160 µV ± 170 µV ± 180 µV
300 kHz
Q
± 4.5 mV
1 V
1 kHz
1 V
200 Hz
± 600 µV ± 800 µV ± 900 µV
± 1 mV
50 kHz
± 1.4 mV ± 1.6 mV ± 1.7 mV ± 1.8 mV
300 kHz
± 45 mV
0.03 V
1 kHz
10 V
200 Hz
± 2 mV
± 3 mV
1 V
1 kHz
200 Hz
± 2.4 mV ± 3.5 mV ± 3.6 mV ± 3.7 mV
10 V
10 Hz
3 Hz
± 6 mV
± 8 mV
± 9 mV
± 10 mV
100 Hz
20 Hz
± 6 mV
± 8 mV
± 9 mV
± 10 mV
20 kHz
200 Hz
± 6 mV
± 8 mV
± 9 mV
± 10 mV
50 kHz
200 Hz
Q
± 14 mV
± 16 mV
± 17 mV
± 18 mV
100 kHz
± 63 mV
± 68 mV
300 kHz
± 450 mV
100 V
1 kHz
100 V
200 Hz
Q
± 60 mV
± 80 mV
± 90 mV ± 100 mV
50 kHz
± 140 mV ± 160 mV ± 170 mV ± 180 mV
70 V
300 kHz
± 3.3 V
750 V
1 kHz
750 V
200 Hz
± 450 mV ± 600 mV ± 675 mV ± 750 mV
210 V
50 kHz
± 510 mV ± 606 mV ± 627 mV ± 648 mV
70 V
300 kHz
± 6.6 V
Q
:
Quick performance verification test point.
Keysight True
volt
Series Operating and Service Guide
517
Performance Verification