CALCulate:SCALe:OFFSet {<
offset
>|MIN|MAX|DEF}
CALCulate:SCALe:OFFSet? [{MIN|MAX|DEF}]
Stores an offset value
B
for the scaling function in
.
Parameter
Typical Return
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to 1.0E+15. Default: 0.0.
+5.0000
Enable
Mx-B
scaling with a gain of 100 and an offset of 5.
CALC:SCAL:FUNC SCAL
CALC:SCAL:GAIN 100
CALC:SCAL:OFFS 5
CALC:SCAL:STAT ON
l
Applies only to the 34465A and 34470A.
l
The instrument sets the offset value to 0.0 after a Factory Reset (
) or Instrument Preset
(
).
CALCulate:SCALe:REFerence {<
reference
>|MIN|MAX|DEF}
CALCulate:SCALe:REFerence? [{MIN|MAX|DEF}]
Stores a reference value for the PCT function in
.
Parameter
Typical Return
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to 1.0E+15. Default: 0.0. +3.0002
Enable PCT scaling with a 100V reference.
CALC:SCAL:REF 100
CALC:SCAL:FUNC PCT
CALC:SCAL:STAT ON
l
Applies only to the 34465A and 34470A.
l
Specifying a reference value disables automatic reference selection (
OFF).
l
The instrument sets the reference value to 0.0 with automatic reference selection enabled after a Fact-
ory Reset (
), or a change in measurement function.
Keysight True
volt
Series Operating and Service Guide
225
SCPI Programming Reference