l
controls whether the unit string is displayed when the scaling function
is enabled.
l
The instrument resets the units to the null string after a Factory Reset (
) or an Instrument Preset
(
).
CALCulate:SCALe:UNIT:STATe {OFF|ON}
CALCulate:SCALe:UNIT:STATe?
Enables or disables displaying the unit string with measurements on the front panel when the scaling func-
tion is enabled.
Parameter
Typical Return
{ON|1|OFF|0}. Default: OFF.
0 (OFF) or 1 (ON)
Enable scaling with a gain of 100 and an offset of 5, and display "PSI" with measurements on the front
panel.
CALC:SCAL:FUNC SCAL
CALC:SCAL:GAIN 100
CALC:SCAL:OFFS 5
CALC:SCAL:UNIT "PSI"
CALC:SCAL:UNIT:STAT ON
CALC:SCAL:STAT ON
l
Applies only to the 34465A and 34470A.
l
specifies the unit string to be displayed.
l
The instrument disables displaying the unit string after a Factory Reset (
) or an Instrument Preset
(
).
SCPI Programming Reference
228
Keysight True
volt
Series Operating and Service Guide