CALCulate:TCHart[:STATe] {ON|1|OFF|0}
CALCulate:TCHart[:STATe]?
Enables or disables trend chart data when controlling the instrument from remote.
Parameter
Typical Return
{OFF|ON|0|1}. Default: OFF.
OFF or ON
Enable trend chart data on 500 measurements:
CALCulate:TCHart:STATe ON
CONF:VOLT:DC 10,0.003
SAMP:COUN 500
INIT
FETC?
Typical Response:
+9.9938382800,+9.9935231500,...
(500 measurements)
l
You must enable this (ON)
before
sending
or
to see trend chart data when controlling the
instrument from remote. If it is disabled or enabled
after
or
, viewing the trend chart res-
ults in a chart with no data.
l
Trend chart data must be disabled (OFF) to make measurements at the highest speed (50 kHz).
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
236
Keysight True
volt
Series Operating and Service Guide