CONFigure Subsystem
The CONFigure commands are the most concise way to configure measurements. Like the
quer-
ies, these commands use default measurement configuration values. However, these commands do not
automatically start measurements, so you can modify measurement attributes before initiating the meas-
urement.
Use
or
to initiate measurements.
Command Summary
CONFigure:{RESistance|FRESistance}
CONFigure[:VOLTage][:DC]:RATio
Default Settings for the CONFigure Command
The CONFigure commands select the function, range and resolution in one command. Specify
<
resolution
> in the measurement's units (V, A, Hz, Ω, and so on). All other parameters are set to their
default values (below).
Measurement Parameter
Default Setting
AC Input Filter (bandwidth)
20 Hz (medium filter)
Autozero
OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
Range
AUTO (including voltage range for frequency and period measurements)
Samples per Trigger
1 sample
Trigger Count
1 trigger
Trigger Delay
Automatic delay
Trigger Source
Immediate
Trigger Slope
NEGative
Math Functions
Disabled. Other parameters are unchanged.
Per-function Null State
Disabled
Keysight True
volt
Series Operating and Service Guide
243
SCPI Programming Reference