SENSe Subsystem Introduction
The SENSe subsystem configures measurements. The most basic SENSe command is
, which selects the measurement function. All other SENSe commands are associated with specific
measurement types:
(34465A and 34470A only)
[SENSe:]FUNCtion[:ON] "<
function
>"
[SENSe:]FUNCtion[:ON]?
Selects the measurement function (all function-related measurement attributes are retained).
Parameter
Typical Return
CAPacitance
CONTinuity
CURRent:AC
CURRent[:DC]
DIODe
FREQuency
FRESistance
PERiod
RESistance
TEMPerature
VOLTage:AC
VOLTage[:DC]
VOLTage[:DC]:RATio
The default is VOLTage
[:DC].
The short form of the selected function is returned in quotation marks, with no
optional keywords:
"CONT", "CURR:AC", "CURR", "DIOD", and so on.
Select the AC voltage function:
FUNC "VOLT:AC"
l
If you change the measurement function, all measurement attributes of the previous function (range,
resolution, etc.) are remembered. If you return to the original function, those measurement attributes
are restored.
l
Changing the measurement function disables scaling, limit testing, histogram, statistics, and trend
chart data collection (
,
and
set to OFF).
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide
313
SCPI Programming Reference