[SENSe:]FREQuency:SECondary {"OFF"|"CALCulate:DATA"|"PERiod"|"VOLTage:AC"}
[SENSe:]FREQuency:SECondary?
Selects the secondary measurement function for frequency measurements.
Parameter
Typical Return
{"OFF"|"CALCulate:DATA"|"PERiod"|"VOLTage:AC"},
default "OFF"
"CALC:DATA"
Configure frequency measurements, select period as the secondary measurement, make and retrieve meas-
urements.
CONF:FREQ MAX, MAX
FREQ:SEC "PERiod"
READ?;DATA2?
Typical Response: +1.2202;+8.18129619E-03
l
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l
“PERiod” - The period measurement of the input signal.
l
"VOLTage:AC" - The AC voltage measurement of the input signal.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
[SENSe:]PERiod:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage:AC"}
[SENSe:]PERiod:SECondary?
Selects the secondary measurement function for period measurements.
Parameter
Typical Return
{"OFF"|"CALCulate:DATA"|"FREQuency"|"VOLTage:AC"},
default "OFF"
"CALC:DATA"
Configure period measurements, select frequency as the secondary measurement, make and retrieve meas-
urements.
CONF:PER MAX, MAX
PER:SEC "FREQ"
READ?;DATA2?
Typical Response: +8.90363477E-03;+1.1202
l
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l
“FREQuency” - The frequency measurement of the input signal.
l
"VOLTage:AC" - The AC voltage measurement of the input signal.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).