l
Applies only to the 34465A and 34470A.
l
For this command, you must always specify the temperature in degrees Celsius regardless of the tem-
perature units currently selected (see
command).
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust
{<
temperature
>|MIN|MAX|DEF}
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:OFFSet:ADJust? [{MIN|MAX|DEF}]
Allows you to make minor temperature adjustments to correct for the differences between the DMM
internal temperature measurement of the front connection and the actual temperature of the meas-
urement terminals.
Parameter
Typical Return
-20 °C to +20 °C. Default: 0 °C.
+2.0001
Configure thermocouple temperature measurements using the internal reference junction. Measure the
temperature inside the instrument. Assuming the temperature outside the instrument is 5 °C cooler than
the inside measurement, adjust the temperature by -5 °C. Then make and read one measurement:
CONF:TEMP TC,K
TEMP:TRAN:TC:RJUN:TYPE INT
TEMP:TRAN:TC:CHEC ON
SYST:TEMP?
TEMP:TRAN:TC:RJUN:OFFS:ADJ -5
READ?
Typical Response: +3.3100
l
Applies only to the 34465A and 34470A.
l
Important:
Since the internal reference temperature is the temperature of the
front
connections, use
of the
rear
connections with the selection of the internal reference junction will have an unknown
error with no specified performance and is not recommended.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
368
Keysight True
volt
Series Operating and Service Guide