l
Applies only to the 34465A and 34470A.
l
Both the [SENSe:]<
function
>:NPLC and [SENSe:]<
function
>:APERture commands control the instru-
ment's integration time. When using both commands (not recommended), whichever command is
sent last will override the other command and set the integration time. Sending the
[SENSe:]<
function
>:APERture command sets the [SENSe:]<
function
>:APERture:ENABled command
to 1 (ON); sending [SENSe:]<
function
>:NPLC sets it to 0 (OFF).
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
[SENSe:]VOLTage[:DC]:APERture:ENABled {ON|1|OFF|0}
[SENSe:]VOLTage[:DC]:APERture:ENABled?
Enables the setting of integration time in seconds (called
aperture time
) for dc voltage measurements. If
aperture time mode is disabled (default), the integration time is set in PLC (power-line cycles).
Parameter
Typical Return
{ON|1|OFF|0}. Default: OFF. 0 (OFF) or 1 (ON)
Enable aperture time mode, set the aperture time to 300 ms:
VOLT:APER:ENAB ON
VOLT:DC:APER 300E-03
l
Applies only to the 34465A and 34470A.
l
The
, and
[SENSe:]VOLTage[:DC]:RESolution
commands disable the aperture time mode, and select an integ-
ration time in number of power line cycles.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
380
Keysight True
volt
Series Operating and Service Guide