SYSTem:LICense:DELete "<
option_name
>"
Deletes a license and disables the associated functionality.
If the security option is installed, you must unsecure the instrument with the security
code (calibration password) to execute this command.
Parameter
Typical Return
"LAN" or "SEC" (34460A)
"SEC" (34461A)
"DIG", "MEM", or "SEC" (34465A/70A)
(none)
Delete LAN license:
SYST:LIC:DEL "LAN"
l
Valid option names are double quoted strings representing the installed
. Installed
options can be easily identified using
l
Executing this command sets the instrument to the
.
SYSTem:LICense:DELete:ALL
Deletes all licenses and disables the associated functionality.
If the security option is installed, you must unsecure the instrument with the security
code (calibration password) to execute this command.
Parameter Typical Return
(none)
(none)
Delete all licenses:
SYST:LIC:DEL:ALL
l
Executing this command sets the instrument to the
.
SYSTem:LICense:DESCription? "<
option_name
>"
Returns a description of the specified option, regardless of whether it is currently licensed.
Parameter
Typical Return
"LAN" or "SEC" (34460A)
"SEC" (34461A)
"DIG", "MEM", or "SEC" (34465A/70A)
"Enable instrument security"
Return description for option 34460A-LAN or option 3446LANU:
SYST:LIC:DESC? "LAN"
l
Valid option names are double quoted strings representing the installed
. Installed
options can be easily identified using
Keysight True
volt
Series Operating and Service Guide
423
SCPI Programming Reference