TRIGger:DELay:AUTO {ON|1|OFF|0}
TRIGger:DELay:AUTO?
Disables or enables automatic trigger delay. If enabled, the instrument determines the delay based on func-
tion, range, and integration time or bandwidth.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON.
0 (OFF) or 1 (ON)
Return 5 DC voltage measurements, with an automatic delay between each measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response:
+4.2700,+4.2700,+4.2700,+4.2700,+4.2700
l
Selecting a specific trigger delay using
disables the automatic trigger delay.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
TRIGger:LEVel {<
level
>|MIN|MAX|DEF}
TRIGger:LEVel? [{MIN|MAX|DEF}]
Sets the level on which a trigger occurs when level triggering is enabled (
set to
INTernal).
It is recommended that all triggered measurements be made using an appropriate fixed
manual range. That is, turn autorange off (
SENSe:<function>:RANGe:AUTO OFF
), or set
a fixed range using the
,
, or
command.
Parameter
Typical Return
<
level
> (see bullet points below). Default: 0.
+2.0001
Configures dc voltage measurements, 1 V range, 0.1 mV resolution. Select internal trigger source. Set the
trigger level to 0.75 volts, positive slope. INIT puts the instrument in the "wait for trigger" mode.
CONF:VOLT:DC 1,0.0001
TRIG:SOUR INT
TRIG:LEV 0.75
TRIG:SLOP POS
INIT
l
Applies only to the 34465A and 34470A, and requires the DIG option.
l
Supported only for ac and dc voltage, ac and dc current, frequency, period, 2-wire resistance, 4-wire
resistance, RTD and thermistor temperature measurements. Resistance and temperature meas-
urements must have offset compensation and low power disabled.
l
To use the trigger level you must select INTernal as the trigger source with the
com-
mand.
SCPI Programming Reference
430
Keysight True
volt
Series Operating and Service Guide