DC Current 10 A Gain Calibration
This calibration procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the
34460A).
Review
and
Flatness Calibration Considerations
before performing any calibration.
Configuration: DC Current
1.
Configure the 10 A range and set the integration time to 100 PLC.
2.
Apply the 5 A input.
3.
Enter the actual applied input current as the Cal Value (see
).A Calibration
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4.
Store the new calibration constants (see
Storing the Calibration Constants
).
This calibration takes less than 4 seconds to complete.