Overview of the 34934A
1
Keysight 34934A User’s Guide
31
connects columns from separate 32-column banks), you can program the mode
for the row protection resistor and bypass relay.
MH
ML
C1L – C32L
C33L – C64L
INST
DUT
-
+
+
-
1503
1535
.
.
.
. . .
.
. .
. . .
1567
1599
ROUT: CLOS: PAIR (@1503 , 1535 )
Paired channels are: 1567 1599
row R3H @ columns C3H and C35H
row R3L @ columns C3L and C35L
R3H
R1H
R2H
R4H
R3L
R1L
R2L
R4L
C1H – C32H
C33H – C64H
Closes:
See “Channel Numbering” on page 168 for the numbering scheme.