Features and Functions
4
Keysight 34970A/34972A User’s Guide
173
Self-test
A
power-on
self-test occurs automatically when you turn on the instrument. This
limited test assures you that the instrument and all installed plug-in modules are
operational. This self-test does not perform the extensive set of tests that are
included as part of the complete self-test described below.
A
complete
self-test runs a series of tests and takes approximately 20 seconds to
execute. If all tests pass, you can have a high confidence that the instrument and
all installed plug-in modules are operational.
– If the complete self-test is successful,
PASS
is displayed on the front panel. If
the self-test fails,
FAIL
is displayed and the
ERROR
annunciator turns on. See
the
34970A/34972A Service Guide
for instructions on returning the instrument
to Keysight for service.
–
Front-Panel Operation
: To perform the complete front-panel self-test, hold
down
as you turn on the instrument and
hold down the key until you
hear a long beep
. The self-test will begin when you release the key following
the beep.
–
Remote Interface Operation:
*TST?
Returns “0” if the self-test is successful or “1” if it fails.