5
Error Messages
236
Keysight 34970A/34972A User’s Guide
-221
Settings conflict
An invalid configuration was requested. This error is most commonly generated
when setting alarm limits. Note that the lower limit must always be less than or
equal to the upper limit, even if you are using only one of the limits. This error is
also generated if you send the
MEASure?
or
CONFigure
command with
autorange enabled with a fixed resolution.
-222
Data out of range
A numeric parameter value is outside the valid range for this command.
Example
:
TRIG:COUNT -3
-223
Too much data
A character string was received but could not be executed because the string
length was more than 12 characters. This error can be generated by the
CAL:STRing and DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example
:
TRIG:SOURCE ALARM
(
ALARM
is not a valid choice)
-230
Data stale
A
FETCh?
or
DATA:REMove?
command was received but internal reading
memory was empty. The readings retrieved may be invalid.
-310
System error
A firmware defect has been found.
This is not a fatal error but you should contact
your nearest Keysight Service Center if this error is reported.
-350
Error queue overflow
The error queue is full because more than 10 errors (34970A) or 20 errors
(34972A) have occurred. No additional errors are stored until you remove errors
from the queue. The error queue is cleared by the
*CLS
(clear status) command
or when power is cycled. The errors are also cleared when you read the queue.
-410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output
buffer contained data from a previous command (the previous data is not
overwritten). The output buffer is cleared when power has been off or after a bus
Device Clear.