4
Features and Functions
106
Keysight 34970A/34972A User’s Guide
– If a scan includes a read of the totalizer, multifunction module, the count is
reset each time it is read during the scan
only
when the totalizer reset mode is
enabled (
TOTalize:TYPE RRESet
command or the
Advanced
menu for the
totalizer).
– If you install a module while a scan is running, the instrument will cycle power
and resume scanning. If you remove a module while a scan is running, the
instrument will cycle power, and it will not resume scanning when the reboot
completes. If you are logging to a USB drive, the instrument will not log any
scans that occur between the removal of the module and the reboot of the
instrument.
– You can use either the internal DMM or an external DMM to make
measurements of your configured channels. However, the instrument allows
only one scan list at a time; you cannot scan some channels using the internal
DMM and others using an external DMM. Readings are stored in 34970A/
34972A memory
only
when the internal DMM is used.
– If the internal DMM is installed and enabled, the instrument will automatically
use it for scanning. For externally-controlled scans, you must either remove
the internal DMM from the 34970A/34972A or disable it (see
“Internal DMM
Disable” on page