Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-17
If the measured dimension does not match the actual, the adjust the Y
Sensitivity term according to the following equation:
Non-Orthogonality
Using the Angle tool, measure the angle between features at all four
corners of the image. The image below (
) shows one of the
four corner measurements.
Figure 8-15
Angle measurement for calibrating Non-orthogonality
Adjust the Non-orthogonality term in the Open Loop tab of the Scanner
Setup dialog box until the angle measures as close to 90 ° as possible for
NewSensitivity
CurrentSensitivit y Kno
wnSize
MeasuredSize
--------------------------------------------------------------------------------------
=