Electrochemistry
5
Keysight 5500 SPM User’s Guide
5-13
•
Bipotentiosat (PicoStat)
•
SECM nosecone with SECM tip
•
Sample plate with EC connection
•
EC cell
The procedure for setting up the components is as follows.
1
Connect AFM Controller as HEB as in normal Contact mode.
2
Connect a SECM module to the microscope base, then connect the
module to HEB with the standard 44-pin cable.
Figure 16-10
SECM Adapter (left) and adapter connected to AFM Base
(right).
3
Assemble the 5500 SECM nose cone (details below).
4
Insert assembled nose cone into scanner.
5
Mount the scanner in the base and align the laser to the cantilever
and detector in air, again, just as in standard Contact mode.
6
Assemble the EC cell as described in
above. Ensure there is enough gap between the tip and
sample plate to avoid damaging the tip when placing the sample
plate in position.
7
Approach tip to sample and run AFM/SECM imaging.