Keysight 5500 SPM Components
5
Keysight 5500 SPM User’s Guide
5-5
Nose Assembly
The nose assembly retains the cantilever and enables its motion. A
spring clip on the nose assembly secures the probe in place. The nose
assembly is held securely in the scanner by an O-ring.
One-Piece Nose Assemblies
The most widely used nose assemblies consist of a single unit which is
installed in the scanner (
). One-piece nose assemblies are
available for different modes and may include additional electronics
and/or components. For example, the Top MAC nose assembly includes
a coil that provides an oscillating magnetic field.
Additionally, nose assemblies are designed to hold the probe at either
nine degrees or eight degrees from horizontal. Nine-degree nose
assemblies are used for general purpose imaging. Nine and eight-degree
nose assemblies may be used for imaging in liquid. A Top MAC nose
assembly is available with an eight-degree nose assembly option.
Figure 2-3
One-piece nose assemblies. Clockwise from upper left: Top
MAC, CSAFM, Contact Mode, AAC, STM
CAUTION
The CSAFM nose assembly cannot be used for imaging in liquid.