Setting Up the Keysight 5500 SPM
5
Keysight 5500 SPM User’s Guide
5-8
Remove the accessory bag from the chamber. This bag contains four
adjustable bungee mounts, four bungee cords and a 2.5 mm hex wrench
for assembly.
Place the instrument platform, included in a smaller box, on the floor of
the chamber.
After unpacking is complete, proceed with assembling the vibration
isolation chamber:
1
Attach bungee mounts to screws mounted on enclosure ceiling by
inserting screw into open end of bungee mount and turning the
mounts' pin to the right.
2
Attach bungee cords to eyebolts on adjustable bungee mount using
bungee cord hook.
3
Lift back side of instrument platform and tilt up towards bungee
hooks. Attach the back two bungee cord hooks, one at a time, to
recessed attachment points as shown in
a
Insert hook into recess hole parallel to the attachment bar.
b
Turn hook ninety degrees.
Figure 3-4
Attaching bungee hook to instrument platform
4
for the front side of the instrument platform.
5
Adjust bungee mounts to ensure that the instrument platform is
resting level. Turn the standoff using the center pin as shown in
CAUTION
Do not adjust bungee mounts with SPM instrument on platform.