Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-2
defined height above the sample for specialized experiments, such as
measuring how long molecules stretch and relax (
Figure 9-1
Setting Z position in the Position tab in the Servo dialog box
X/Y/Z Closed-Loop Scanner
The X/Y/Z closed-loop scanner includes the functionality of the Z-axis
scanner described above, with additional encoders on the X and Y axes.
These sensors allow for very linear scans and also make it easy to move
to precise locations on the sample.
Calibration
Keysight recommends calibrating the Gains, Offset and Sensitivity
settings in the Closed Loop tab of the Scanner Setup dialog box prior to
each use of the closed-loop scanner. Begin by calibrating the scanner
following the open-loop instructions in the
For a Z-axis scanner, skip to
NOTE
The procedures and descriptions in this User Guide are for PicoView
1.12x and higher.