Electrochemistry
5
Keysight 5500 SPM User’s Guide
5-14
Assembling the SECM Nose Cone
The SECM probe comes pre-mounted on a cartridge for easy placement
into the 5500 nose cone.
Figure 16-11
SECM nose cone and probe cartridge
The procedure to assembly the SECM nose cone is as follows.
1
Strip the Teflon from the last 1/8” of the thin platinum (Pt) wire that
is attached to the probe. This enables the electrical contact between
the probe and the EC electronics.
2
Carefully hold the cartridge and slide it into the nose cone, aligning
the platform holding the probe with the slot in the nose cone as seen
in