Cleaning Instructions
Which Cleaning Procedure should I use ?
82
Keysight 81630B/4B/6B Power Sensor Modules and Keysight 81635A Dual Power Sensor Module, Edition 4
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an
ultrasonic bath with isopropyl alcohol.
1
Hold the tip of the connector in the bath for at least three minutes.
2
Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
3
Blow away any remaining lint with compressed air.