20
Keysight 87406Q Operating and Service Manual
3
Specifications
General Specifications
Specifications refer to the performance standards or limits against
which the switch is tested.
Typical characteristics are included for additional information only
and they are not specifications. These are denoted as "typical",
"nominal", or “approximate" and are printed in italics.
Maximum power rating
Into internal termination
1 W CW
50 W peak, 10
μ
s max pulse width, not to exceed
1 W average
Into thru path
Hot switching
2 W CW
100 W peak, 10
μ
s max pulse width, not to exceed
2 W average
Cold switching
See
“Supplemental specifications (cold switching)”
Switching time
15 ms maximum