2-6
Keysight B1500A Configuration and Connection Guide, Edition 6
B1500A Accessories
Accessories for the B1500A
2.1.3 HVSPGU accessories
•
SPGU cables
- have co-axial type (SMA to coaxial) cables.
•
SPGU synchronous cables
- have co-axial (SMA to SMA) type cables. If multiple
SPGUs have been installed, connect the SPGUs by this cable. This connection makes it
possible to perform the synchronous output by the multiple SPGU channels. The
master SPGU is the module which outputs the reference signal and the slave SPGUs
are the modules which follow to the reference signal.
•
SMU/PGU selector
- Used to switch the measurement resource connected to the DUT.
The measurement resource can be SPGU or SMU. Needs the 16445A adapter to use the
selector. One selector provides two switching channels.
•
Digital I/O cables
- General digital I/O connection cable. This cable can connect
between digital I/O port on the B1500A and 16445A SMU/PGU selector adapter.
Table 2-3
HVSPGU accessories for B1500A semiconductor device analyzer
Model/Option
Description
Additional Information
SPGU cable
16493P-001
SPGU cable (SMA to BNC), 1.5 m
16493P-002
SPGU cable (SMA to BNC), 3 m
SPGU synchronous cable
16493Q-001
SPGU synchronous cable (SMA to SMA), 8 cm
SMU/PGU selector
16440A
SMU PGU selector
One 16440A has 2 channels and each channel can
switch an SMU and a pulse generator.
16440A-003
Control cable, 40 cm
Connect between 16440A and 16445A