Triggers
11
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
181
OR Trigger
The OR trigger mode triggers when any one (or more) of the specified edges on
analog or digital channels is found.
1
On the front panel, in the Trigger section, press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey, and use the Entry knob to select
OR
.
3
Press the
Slope
softkey and select rising edge, falling edge, either edge, or don't
care. The selected slope is displayed in the upper-right corner of the display.
4
For each analog or digital channel you want to include in the OR trigger, press
the
Channel
softkey to select the channel.
As you press the
Channel
softkey (or rotate the Entry knob), the channel you
select is highlighted in the OR = line directly above the softkeys and in the
upper-right corner of the display next to the OR gate symbol.
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob. Press the
[Digital]
key and select
Thresholds
to set the threshold level
for digital channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
5
For each channel you select, press the
Slope
softkey and select (Rising),
(Falling), (Either), or X (Don't Care). The selected slope is displayed above the
softkeys.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...