Acquisition Control
13
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
231
Measurements, Statistics, and Infinite Persistence with Segmented
Memory
To perform measurements and view statistical information, press
[Meas]
and set up
your desired measurements (see
, “Measurements,” starting on page
243). Then, press
Analyze Segments
. Statistical data will be accumulated for the
measurements you have chosen.
The
Analyze Segments
softkey appears when the acquisition is stopped and the
segmented memory feature is on or when the serial Lister is enabled.
You can also turn on infinite persistence (in the Display Menu) and press the
Analyze Segments
softkey to create an infinite persistence display.
Segmented Memory Re-Arm Time
After each segment fills, the oscilloscope re-arms and is ready to trigger in about
1 µs.
Remember though, for example: if the horizontal time per division control is set to
5 µs/div, and the Time Reference is set to
Center
, it will take at least 50 µs to fill all
ten divisions and re-arm. (That is 25 µs to capture pre-trigger data and 25 µs to
capture post-trigger data.)
Saving Data from Segmented Memory
You can save either the currently displayed segment (
Save Segment - Current
), or all
segments (
Save Segment - All
) in the following data formats: CSV, ASCII XY, and
BIN.
Be sure to set the Length control to capture enough points to accurately represent
the captured data. When the oscilloscope is busy saving multiple segments,
progress is displayed in the upper right area of the display.
For more information, see
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...