Measurements
15
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
271
•
The units for absolute threshold for each channel is set in the channel probe
menu.
4
Press the
Lower
softkey; then, turn the Entry knob to set the lower measurement
threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from 0% to
98%.
5
Press the
Middle
softkey; then, turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper thresholds.
The default middle threshold is 50% or 1.20 V.
•
If threshold
Type
is set to
%
, the middle threshold value can be set from 1% to
99%.
6
Press the
Upper
softkey; then, turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will automatically
decrease the middle value to be less than the upper value. The default upper
threshold is 90% or 1.50 V.
•
If threshold
Type
is set to
%
, the upper threshold value can be set from 2% to
100%.
TIP
Absolute threshold hints
•
Absolute thresholds are dependent on channel scaling, probe attenuation, and probe units. Always
set these values first before setting absolute thresholds.
•
The minimum and maximum threshold values are limited to on-screen values.
•
If any of the absolute threshold values are above or below the minimum or maximum waveform
values, the measurement may not be valid.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...