Trigger Mode/Coupling
12
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
213
Changing the time base settings does not affect the trigger holdoff time.
With Keysight's MegaZoom technology, you can press
[Stop]
, then pan and zoom
through the data to find where the waveform repeats. Measure this time using
cursors; then, set the holdoff.
External Trigger Input
The external trigger input can be used as a source in several of the trigger types.
The external trigger BNC input is labeled
EXT TRIG IN
.
The external trigger input impedance is 1M Ohm. This lets you use passive probes
for general-purpose measurements. The higher impedance minimizes the loading
effect of the oscilloscope on the device under test.
To set the EXT TRIG IN units and probe attenuation:
1
Press the
[Mode/Coupling]
key in the Trigger section of the front panel.
2
In the Trigger Mode and Coupling Menu, press the
External
softkey.
3
In the External Trigger Menu, press the
Units
softkey to select between:
•
Volts
— for a voltage probe.
•
Amps
— for a current probe.
CAUTION
Maximum voltage at oscilloscope external trigger input
300 Vrms, 400 Vpk
1M ohm input: For steady-state sinusoidal waveforms derate at 20 dB/decade above
100 kHz to a minimum of 5 Vpk
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...