Keysight E1412A Digital Multimeter User Guide
cxxix
MEASure
Multimeter Command Reference
–
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making DC Current Measurements
MEAS:CURR:DC? .1,MAX
Function: DC current; range
selected: 1A MAX resolution: 1.0E-05A.
:FREQuency?
MEASure:FREQuency? [<
range
>|MIN|MAX|DEF|AUTO
[,<
resolution
>|MIN|MAX|DEF]]
selects the frequency function and uses one range
for all inputs between 3Hz and 300kHz.
Parameters
Comments
–
The frequency function uses one “range” for all inputs between 3Hz and
300kHz. Querying the range will always return “3E+00”. A frequency
measurement returns “
0
” if no input is applied.
–
Range and resolution settings are listed below for the
MIN
,
MAX
,
DEF
and
AUTO
parameters and after a module reset (
*RST
).
:FRESistance?
MEASure:FRESistance? [<
range
>|MIN|MAX|DEF|AUTO
[,<
resolution
>|MIN|MAX|DEF]]
selects the 4-wire ohms function and allows you
to specify the measurement range and resolution.
Parameter Name Parameter Type
Range of Values
Default
Units
<
range
>
numeric
3E+00
Hz
<
resolution
>
numeric
3E-04 | 3E-05 | 3E-06
Hz
PARAMETER
RANGE
RESOLUTION
MIN
3E+00
3E+06
MAX
3E+00
3E+04
DEF|AUTO
and module
reset (
*RST)
3E+00
3E+05