cxxxvi
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
MEASure
–
To specify a
MIN
or
MAX
resolution while autoranging, you must specify
MEAS:VOLT:DC:RAT? DEF
(you cannot omit the
range
parameter). This
prevents the
MIN
or
MAX
resolution from being interpreted as a range
setting and the resulting command becomes
MEAS:VOLT:DC:RAT? DEF,MIN
or
MEAS:VOLT:DC:RAT? DEF,MAX
.
–
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making DC Voltage Ratio Measurements
MEAS:VOLT:DC:RAT? 0.825,MAX
Function: DC voltage; range
selected: 1V; MAX resolution: 100
V.