[SENSe:]
Multimeter Command Reference
–
The present current range selected if
MIN
or
MAX
is not specified. Only the
ranges available with the
RANGe
command are returned. For example, if
CONFigure
sets the 3A range, 3A is the range returned.
–
The minimum current range available (1A) if
MIN
is specified.
–
The maximum current range available (3A) if
MAX
is specified.
Example
Query the AC Current Measurement Range
CURR:AC:RANG 3
Select 3A range.
CURR:AC:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
CURRent:AC:RANGe:AUTO
[SENSe:]CURRent:AC:RANGe:AUTO <
mode
>
enables or disables the autorange
function for AC current measurements.
Parameters
Comments
–
You can substitute decimal values for the OFF (“0”) and ON (“1”)
parameters.
–
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
–
If you explicitly select a range using CURRent:AC:RANGe, autoranging is
turned OFF.
–
Related Commands: CONFigure, :DC:RANGe, RESistance:RANGe
–
*RST Condition: CURR:AC:RANG:AUTO ON
Example
Disable AC Current Autoranging
CURR:AC:RANG:AUTO OFF
Disable autorange.
Parameter Name Parameter Type
Range of Values
Default
Units
<
mode
>
boolean
OFF|0|ON|1
none