[SENSe:]
Multimeter Command Reference
–
*RST
Condition: 10 PLC
Example
Set the DC Current Integration Time in PLCs
CURR:DC:NPLC 100
Integration time is 100 PLCs.
CURRent[:DC]:NPLC?
[SENSe:]CURRent[:DC]:NPLCycles? [MIN|MAX]
returns one of the following
numbers to the output buffer:
–
The present integration time in PLCs if
MINimum
or
MAXimum
is not
specified.
–
The minimum integration time available (0.02) if
MIN
is specified.
–
The maximum integration time available (100) if
MAX
is specified.
Example
Query the DC Current Integration Time
CURR:DC:NPLC 100
Integration time is 100 PLCs.
CURR:DC:NPLC?
Query multimeter to return integration time.
enter statement
Enter value into computer.
CURRent[:DC]:RANGe
[SENSe:]CURRent[:DC]:RANGe <
range
>
selects the range for DC current
measurements.
Parameters
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected current. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
CURRent:DC:RANGe
command: 10mA.
MAX
selects the maximum range available: 3A.
–
You must select a range using
CURRent:DC:RANGe
before specifying
resolution.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
0.01A|0.1A|1A|3A|MIN|M
AX
amps