clii
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
–
Specifying a fixed range disables the autorange mode set by the
CURR:DC:RANG:AUTO
command.
–
The
CURR:DC:RANG
command overrides the range setting from a previous
CONFigure
command on the same function.
–
*RST
Condition:
CURR:DC:RANG 1
Example
Set the DC Current Range to 3A
CURR:DC:RANG 3
DC Current range is 3A.
CURRent[:DC]:RANGe?
[SENSe:]CURRent[:DC]:RANGe? [MIN|MAX]
returns one of the following
numbers to the output buffer:
–
The present current range selected if
MIN
or
MAX
is not specified. Only the
ranges available with the
RANGe
command are returned. For example, if
CONFigure
selects the 100mA range, 100mA is the range returned.
–
The minimum current range available (10mA) if
MIN
is specified.
–
The maximum current range available (3A) if
MAX
is specified.
Example
Query the DC Current Measurement Range
CURR:DC:RANG 3
Select 3A range.
CURR:DC:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
CURRent[:DC]:RANGe:AUTO
[SENSe:]CURRent[:DC]:RANGe:AUTO <
mode
>
enables or disables the
autorange function for DC current measurements.
Parameters
–
Comments
–
You can substitute decimal values for the
OFF
(“
0
”) and
ON
(“
1
”)
parameters.
Parameter Name Parameter Type
Range of Values
Default
Units
<
mode
>
boolean
OFF|0|ON|1
none