Multimeter Command Reference
–
Specifying a parameter less than 200 but greater than 20 selects the 20Hz
filter.
–
Specifying a parameter less than 20 but greater than 3 selects the 3Hz
filter.
–
Specifying a parameter greater than 200 but not greater than 300E+03
(300kHz) selects the 200Hz filter.
–
Any value greater than 300kHz will cause a “Data out of range” error. The
maximum range for all three filters is 300kHz. Specify the lowest frequency
expected in the input signal. The multimeter selects the appropriate filter
based on the table below.
Comments
–
The following table lists the filter frequency range and the settling time for
making measurements.
–
The ac filter selection is stored in volatile memory and returns to the 20Hz
filter (medium) when power is removed or after a module reset. The upper
limit on all three filters is 300kHz.
–
The slow filter is 3Hz, the medium filter is 20Hz and the fast filter is
200Hz.
–
The
CONFigure
and
MEASure
commands select the 20Hz filter.
–
*RST
Condition:
DET:BAND 20
(medium filter)
Example
Set the ac Signal Filter for Fast Measurements From 200Hz to 300kHz
DET:BAND 200
Selects the fast filter.
DETector:BANDwidth?
[SENSe:]DETector:BANDwidth?
returns which ac filter has been selected. The
value returned is
3
,
20
or
200
. The value is sent to the output buffer.
AC Voltage or Current
Input Frequency
AC Filter
Selected
Max Reading Rate for
Adequate Settling
3Hz to 300kHz
Slow filter
1 reading/7 seconds
20Hz to 300kHz
Medium filter
(default)
1 reading/second
200Hz to 300kHz
Fast filter
10 readings/second