Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
FRES:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
FRESistance:NPLC
[SENSe:]FRESistance:NPLCycles <
number
>
sets the integration time in number
of power line cycles (NPLCs). The NPLC is set to a value from the range of values
that can accommodate the
<
number
>
you specify. For example, specifying
9
sets
the NPLC to 10; specifying
11
sets the NPLC to 100.
Parameters
Comments
–
MINimum
selects 0.02 PLCs.
MAXimum
selects 100 PLCs. Setting the
integration time in power line cycles (PLCs) also sets the aperture time and
the resolution. For example, 10 PLCs sets an aperture time of 167ms (60Hz
line frequency) or 200ms (50Hz). The corresponding resolution depends on
the function and range you select.
–
The
FRES:NPLC
command overrides the results of previously executed
FRESistance:APERture
and
FRESistance:RESolution
commands (the last
command executed has priority).
–
The greater the number of PLCs, the greater the normal mode rejection
(and the lower the reading rate).
–
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode
rejection of 50Hz or 60Hz power line related noise. The 0.02 and 0.2
fractional PLC settings do not provide normal mode rejection of power line
related noise.
–
*RST
Condition:
10 PLC
Example
Set the Integration Time in PLCs
FRES:NPLC 100
Integration time is 100 PLCs.
FRESistance:NPLC?
[SENSe:]FRESistance:NPLC? [MIN|MAX]
returns one of the following numbers
to the output buffer:
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.02|0.2|1|10|100|MIN|MAX
PLCs