Keysight E1412A Digital Multimeter User Guide
clxi
[SENSe:]
Multimeter Command Reference
–
The present integration time in PLCs if
MINimum
or
MAXimum
is not
specified.
–
The minimum integration time available (0.02) if
MIN
is specified.
–
The maximum integration time available (100) if
MAX
is specified.
Example
Query the Integration Time
FRES:NPLC 100
Integration time is 100 PLCs.
FRES:NPLC?
Query multimeter to return integration time.
enter statement
Enter value into computer.
FRESistance:RANGe
[SENSe:]FRESistance:RANGe <
range
>
selects the range for 4-wire resistance
measurements.
Parameters
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected resistance. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
FRESistance:RANGe
command: 100
.
MAX
selects the maximum range available: 100
.
–
You must select a range using
FRESistance:RANGe
before specifying
resolution.
–
Specifying a fixed range disables the autorange mode set by the
FRES:RANG:AUTO
command.
–
The
FRES:RANG
command overrides the range setting from a previous
CONFigure
command on the same function. The multimeter uses the same
aperture time to set the resolution on the new range as was selected by
CONFigure
.
–
*RST
Condition:
FRES:RANG 1
k
(1E+03)
Parameter Name
Parameter Type
Range of Values
Default
Units
<
range
>
numeric
100
|1k
|10k
|100k
|1M
|10M
|100M
|MIN|MAX
ohms