Keysight E1412A Digital Multimeter User Guide
clxiii
[SENSe:]
Multimeter Command Reference
FRES:RANG:AUTO ON
Autorange is turned on for 4-wire ohms
measurements.
FRESistance:RANGe:AUTO?
[SENSe:]FRESistance:RANGe:AUTO?
returns a number to show whether the
autorange mode is enabled or disabled: “
1
” =
ON
, “
0
” =
OFF
. The number is sent
to the output buffer.
Example
Query the Autorange Mode
FRES:RANG:AUTO OFF
Disable autorange.
FRES:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
FRESistance:RESolution
[SENSe:]FRESistance:RESolution <
resolution
>
selects the resolution for 4-wire
resistance measurements.
Parameters
Comments
–
MINimum
selects the best resolution (the smallest value) for the selected
range.
MAXimum
selects the worst resolution (the largest value) for the
selected range.
–
You must select a range using
FRESistance:RANGe
before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the
range selected during autoranging.
–
Specify resolution in the same units as the measurement function.
–
If autoranging is required, set the resolution using the
MIN
or
MAX
parameters or select a specific integration time using
FRESistance:NPLCycles
.
–
*RST
Condition:
FRES:RES 1m
(1E-03)
Parameter Name
Parameter Type
Range of Values
Default
Units
<
resolution
>
numeric
resolution
|MIN|MAX
ohms