clxiv
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
FRESistance:RESolution?
[SENSe:]FRESistance:RESolution? [MIN|MAX]
returns one of the following
numbers to the output buffer:
–
The present resolution selected if
MIN
or
MAX
are not specified. Only the
resolution values available on ranges set by the
RANGe
command are
returned.
–
The resolution with the smallest value (i.e., the best resolution) for the
selected range if
MIN
is specified.
–
The resolution with the largest value (i.e., the worst resolution) for the
selected range if
MAX
is specified.
Example
Query the Resolution
FRES:RES 0.3E-03
Set resolution to 0.3m
.
FRES:RES?
Query multimeter to return the present
resolution.
enter statement
Enter value into computer.
PERiod:APERture
[SENSe:]PERiod:APERture <
time
>|MIN|MAX
sets the integration time in
seconds. Values for time are rounded up to the nearest aperture time shown in
the following table.
Parameters
Comments
MINimum
sets the aperture time to 10ms.
MAXimum
sets the aperture time to
1 second.
The fastest aperture time available when autoranging is 100ms. In order to
specify an aperture time of 10ms, you must select a fixed range.
Setting the aperture time also sets the resolution. Aperture time of 0.01 sets
resolution at 4½-digits, 0.1 sets 5½-digits and 1 sets 6½-digits.
*RST
Condition: 0.1 (100ms)
Parameter Name
Parameter Type
Range of Values
Default
Units
<
time
>
numeric
10ms|100ms|MIN|MAX
seconds