Keysight E1412A Digital Multimeter User Guide
clxxi
[SENSe:]
Multimeter Command Reference
–
The minimum resistance range available (100
) if
MIN
is specified.
–
The maximum resistance range available (100M
) if
MAX
is specified.
Example
Query the Measurement Range
RES:RANG 100
Select 100
range.
RES:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
RESistance:RANGe:AUTO
[SENSe:]RESistance:RANGe:AUTO <
mode
>
enables or disables the autorange
function for resistance measurements.
Parameters
Comments
–
You can substitute decimal values for the
OFF
(“
0
”) and
ON
(“
1
”)
parameters.
–
When autoranging is
ON
, the multimeter samples the input before each
measurement and selects the appropriate range.
–
If you explicitly select a range using
RESistance:RANGe
, autoranging is
turned
OFF
.
–
Related Commands:
CONFigure
,
RESistance:RANGe
–
*RST
Condition:
RES:RANG:AUTO ON
Example
Disable Autoranging
RES:RANG:AUTO OFF
Disable autorange.
RESistance:RANGe:AUTO?
[SENSe:]RESistance:RANGe:AUTO?
returns a number to show whether the
autorange mode is enabled or disabled: “
1
” =
ON
, “
0
” =
OFF
. The number is sent
to the output buffer.
Parameter Name Parameter Type
Range of Values
Default
Units
<
mode
>
boolean
OFF|0|ON|1
none