cxcii
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
TRIGger
Comments
–
MIN
selects the minimum delay of 0 seconds for all functions.
MAX
selects
the maximum delay of 3600 seconds for all functions.
–
The trigger delay is inserted between the trigger and each measurement.
–
If a trigger delay is specified using the
TRIG:DEL <
period
>
,
TRIGger:DELay:AUTO
is turned
OFF
.
–
The multimeter selects an automatic delay if you do not specify a trigger
delay (see the
TRIGger:DELay:AUTO
command).
–
*RST
Condition:
TRIGger:DELay:AUTO ON
Example
Set the Trigger Delay
TRIG:DEL .002
Wait 2ms between trigger and start of measurement.
:DELay?
TRIGger:DELay? [MIN|MAX]
returns one of the following numbers to the output
buffer:
–
The present trigger delay (0 through 3600 seconds) if
MIN
or
MAX
is not
specified.
–
The minimum trigger delay available (0 seconds) if
MIN
is specified.
The maximum trigger delay available (3600 seconds) if
MAX
is specified.
Example
Query the Trigger Delay
TRIG:DEL .002
Wait 2ms between trigger and start of
measurement.
TRIG:DEL?
Query multimeter to return trigger delay.
enter statement
Enter value into computer.
:DELay:AUTO
TRIGger:DELay:AUTO <
mode
>
enables or disables a trigger delay automatically
determined by the present function, range, NPLC setting, AC filter setting and
integration time (see the table on the next page). The trigger delay specifies the
period between the trigger signal and the start of the measurement (and
between each sample when
SAMPle:COUNt > 1
).
Summary of Contents for E1412A