Keysight E1412A Digital Multimeter User Guide
cci
IEEE 488.2 Common Command Quick Reference
Multimeter Command Reference
value.
<
unmask
>
is the sum of the decimal weights of the bits to be enabled
allowing these bits to pass through to the summary bit RQS (bit 6 in the status
byte).
*SRE?
returns the current enable unmask value.
Parameters
A 1 in a bit position enables service request generation when the corresponding
Status Byte Register bit is set; a 0 disables it.
–
Comments
–
Executable when Initiated: Yes
–
Coupled Command: No
–
*RST
Condition: unaffected
–
Power-On Condition: no bits are enabled
Example
Enable Service Request on Message Available Bit
*SRE 16
The firmware revision field will change whenever the firmware is revised.
A.01.00-A.01.00 is the initial revision. The first two digits indicate the major revision
number and increment when functional changes are made. The last two digits indicate
the functional improvement level.
Enable request on MAV.
*STB?
*STB?
returns the value of the Status Byte Register. The RQS bit (bit 6 in the
status byte having decimal weight 64) is set if a service request is pending.
Comments
–
Executable when Initiated: Yes
–
Coupled Command: No
–
Related Commands:
*SRE
–
*RST
Condition: none
Parameter Name
Parameter Type
Range of Values
Default
Units
<
unmask
>
numeric
0 through 255
none