ccxxxiv
Keysight E1412A Digital Multimer User Guide
Measurement Speed and Accuracy Trade-offs
Speed Advantage Using the Special Non-SCPI Commands
(F1-F4 and R1-R7)
You can save approximately three (3) milliseconds by using an
F1
-
F4
special
function command instead of changing function with the equivalent SCPI
[SENSe:]
function command. You can save approximately five (5) milliseconds by
using an
R1
-
R7
special range command instead of changing the range with the
equivalent SCPI
[SENSe:]
range command.
Special Commands and Their Equivalent [SENSe:] Command
Speed Advantage
Special
Command
Equivalent [SENSe:] SCPI command
approximate 3 mS
advantage with
special command
F1
[SENSe:]FUNC "VOLT[:DC]"
F2
[SENSe:]FUNC "CURR[:DC]"
F3
[SENSe:]FUNC "RES"
F4
[SENSe:]FUNC "FRES"
The special range command affects the currently configured function.
approximate 5 mS
advantage with
special command
R1
[SENSe:]VOLT[:DC]:RANG 0.1
[SENSe:]CURR[:DC]:RANG 0.01
[SENSe:]RES:RANG 100
[SENSe:]FRES:RANG 100
R2
[SENSe:]VOLT[:DC]:RANG 1
[SENSe:]CURR[:DC]:RANG 0.1
[SENSe:]RES:RANG 1000
[SENSe:]FRES:RANG 1000
R3
[SENSe:]VOLT[:DC]:RANG 10
[SENSe:]CURR[:DC]:RANG 1
[SENSe:]RES:RANG 10000
[SENSe:]FRES:RANG 10000
R4
[SENSe:]VOLT[:DC]:RANG 100
[SENSe:]CURR[:DC]:RANG 3
[SENSe:]RES:RANG 1E5
[SENSe:]FRES:RANG 1E5
R5
[SENSe:]VOLT[:DC]:RANG 300
CURR[:DC] DOES NOT APPLY
[SENSe:]RES:RANG 1E6
[SENSe:]FRES:RANG 1E6
R6
VOLT[:DC] DOES NOT APPLY
CURR[:DC] DOES NOT APPLY
[SENSe:]RES:RANG 10E6
[SENSe:]FRES:RANG 10E6
R7
VOLT[:DC] DOES NOT APPLY
CURR[:DC] DOES NOT APPLY
[SENSe:]RES:RANG 100E6
[SENSe:]FRES:RANG 100E6
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter User Manual and SCPI Programming Guide 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 18: ...xviii Keysight E1412A Multimeter User Guide ...
Page 242: ......