Keysight E1412A Digital Multimeter User Guide
xlix
Measurement Configuration
Keysight E1412A Multimeter Application Information
–
The multimeter has autorange mode enabled at power-on and after a
module reset.
–
Autorange thresholds:
Down range at <10% of range.
Up range at >120% of range.
–
The multimeter will provide an overload indication by returning
"9.9037"
if the input signal is greater than the present range
can measure and autoranging is disabled or at the maximum range
setting.
–
The multimeter uses one “range” for all inputs between 3Hz and 300kHz
for the frequency and period functions. The multimeter determines an
internal resolution based on a 3Hz signal. If you query the range, the
multimeter will respond with
"3Hz"
. Frequency and period measurements
return
"0"
with no input signal applied.
–
The specified range applies to the signal connected to the Input terminals
for ratio measurements. Autoranging is automatically selected for
reference voltage measurements on the Sense terminals.
–
You can set the range using any of the following commands:
CONFigure:<
function
> <
range
>|MIN|MAX|DEF,<
resolution
>|MIN|MAX|DEF
MEASure:<
function
>? <
range
>|MIN|MAX|DEF,<
resolution
>|MIN|MAX|DEF
[SENSe:]<
function
>:RANGe <
range
>|MIN|MAX
[SENSe:]<
function
>:RANGe:AUTO OFF|ON