Triggering the Multimeter
The Trigger Source
The
TRIGger:SOURce <
source
>
command configures the multimeter's trigger
system to respond to the specified source. The following trigger sources are
available:
–
BUS:
Trigger source is the GPIB Group Execute Trigger (GET) or the
*TRG
common command. Within the 75000 Series C mainframes, the instrument
whose trigger source is “
BUS
” and was the last instrument addressed to
listen will respond to the GPIB Group Execute Trigger. The
*TRG
command
differs from GET because it is sent to a specific instrument not a group of
instruments. NOTE: B-size controllers do not support the BUS trigger (e.g.,
E1306A command module, E1300/E1301A B-size mainframes).
–
EXTernal:
Trigger source is the multimeter's external trigger BNC
connector (labeled “Trig” on the front panel). A falling (negative-going)
edge of the input signal triggers the multimeter. The external pulse signal
must be >1
s, +5V maximum to 0V (TTL levels).
–
IMMediate:
Internal trigger is always present. If the multimeter is in the
wait-for-trigger state (
INITiate
),
TRIGger:SOURce IMMediate
sends the
trigger. The
MEASure
and
CONFigure
commands automatically set the
trigger source to
IMMediate
.
–
TTLTrg0
through
TTLTrg7:
Trigger source is the VXIbus TTL trigger lines.
The multimeter is triggered on the falling (negative- going) edge of a TTL
input signal. NOTE: B-size controllers do not support VXIbus TTL triggers
(e.g., E1306A Command Module, E1300/E1301B B-Size Mainframes).
For example, the following program statement selects the external trigger BNC
connector as the trigger source.
TRIGger:SOURce EXTernal
You can change the trigger source only when the multimeter is in the idle state.
Attempting to change the trigger source while the multimeter is in the
wait-for-trigger state will generate the “Settings conflict” error.