Keysight E1412A Digital Multimeter User Guide
lix
Triggering the Multimeter
Keysight E1412A Multimeter Application Information
The Trigger Count
The
TRIGger:COUNt <
number
>
command sets the number of triggers the
multimeter will accept in the wait-for-trigger state before returning to the idle
state. Use the
number
parameter to set the trigger count to a value between 1
and 50,000. The
MEASure
and
CONFigure
commands set trigger count to 1.
Substituting
MIN
for the
number
parameter sets the trigger count to 1.
Substituting
MAX
for the
number
parameter sets the trigger count to 50,000.
Example: Setting the Trigger Count
In the following example, one DC voltage measurement is made each time the
multimeter's external trigger BNC connector is pulsed low. After 10 external
triggers are received, the multimeter returns to the idle state.
dimension array
Dimension computer array.
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is external BNC on multimeter front panel.
TRIG:COUN 10
Multimeter will accept 10 external triggers (one measurement is
taken per trigger).
READ?
Place multimeter in wait-for-trigger state; make measurements when
external trigger is received; send readings to output buffer.
timeout may occur
May require INIT, monitor the status byte for completion (standard
event bit 0), FETC? to transfer readings to the output buffer (vs.
READ?).
enter statement
Enter readings into computer.
Checking the Trigger Count
The
TRIGger:COUNt? [MINimum|MAXimum]
command returns one of the
following numbers to the output buffer:
–
The present trigger count (1 through 50,000) if neither
MIN
nor
MAX
is
specified.
–
The minimum trigger count available (1) if
MIN
is specified.
–
The maximum trigger count available (50,000) if
MAX
is specified.